Practical-instrument-electronics 830 Manual do Utilizador Página 19

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 52
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 18
Page 17
T/C
All mA∙ Read V
-
23
RTD
RTD
65
4
+
Source & Read Hz mV
Source V∙ pH
-
+
1
-
All milliamp connections
To
Sensor
To
4-20 mA Loop
Typical
2-Wire
Transmitter
(Disconnected)
+ IN - REF +OUT-
Receiver
Powers External
2-Wire Transmitter
Power Supply
(2 to 60 VDC)
500.0
°C
12.000mA
27.4 V
830 Ω
5 6
+
-
Connecting the 830 in place of a transmitter
Connect the PIE 830 in place of the transmitter. The 4-20 mA
loop connects to the mA jacks of the 830 and the sensor (or
pressure module) connects to the other jacks of the 830.
Note: With LOOPSCOPE on the 830 occasionally performs a
test on the loop. During this short test the loop mA signal will be
affected.