Page 15
READ
%
PWR MEASURE
28.20
SOURCE
HI
SET
LO
OFF
MAX
READ
MIN
READ
OVERLOAD
Hz SYNC
Double Click
Menu
Push & Turn
for
Fast Dialing
Push & Hold
to
Store/Step
mA ∙ V ∙ pH ∙ TC ∙ Ω ∙ RTD ∙ Freq ∙ Pressure
Typical
2-Wire
Transmitter
+ IN - REF +OUT-
Transmitter Input
Sensor
Process Signal
Simulated Input
SOURCE
HI
SET
LO
MAX
READ
MIN
READ
Power/Measure mA, Power/Measure % (Percent of 4 to 20 mA)
Choose this function to simultaneously supply power to a 2 Wire
Transmitter while displaying the 4.000 to 20.000 mA output of
the transmitter.
Move the power switch
w to READ then Double Click the
EZ-DIAL knob to get into the Double Click Menu. Turn the knob
e to scroll through the settings and press the knob to make your
selection. Select mA for the FUNCTION and PWR MEAS for the
MODE. Choose either mA or % and whether you need the 250Ω
HART resistor active in the loop.
Disconnect one or both input wires from the device to be
calibrated. Connect the red source lead of the PIE 820-ELITE to
the plus (+) input of the device and the black source lead to the
minus (-).
The PIE 820-ELITE supplies a nominal 24 volts DC at 24 mA to
the 2 Wire Transmitter. The current passed by the transmitter will
be accurately displayed by the PIE 820-ELITE. Calibrate the
transmitter in the usual manner and disconnect the PIE
820-ELITE. Signals above 24 mA are current limited by protection
circuitry with “OVERRANGE” flashed on the display and the red
OVERLOAD LED lit.