Connection Diagram
Two Wire Connection to Transmitter
Specifications
General Specifications:
Unless otherwise indicated all specifications are rated from a nominal 23 °C, 70 % RH for 1 year from calibration.
Temperature Range -25 to 60 °C (-10 to 140 °F)
Relative Humidity Range 10 % ≤RH ≤90 % (0 to 35 °C), Non-condensing
10 % ≤RH≤ 70 % (35 to 60 °C), Non-condensing
4.9 X 3.15 X 1.82 inches (125.5 X 80 X 46.2 mm)
Overall Weight (including 9V battery) 7.2 oz (204 grams)
Battery 9V Alkaline provides 45 hours of continuous use
Low battery indication with nominal 1 hour of operation left
Overload protected to 60 volts for 30 seconds or less
High-contrast graphic liquid crystal display with 0.357” (9.07 mm) high digits
Millivolt Accuracy ±(0.008 % of mV Setting + 0.006 mV)
Temperature Coefficient of mV Source
50 ppm/°C of output range
Cold Junction Calibration Accuracy ±0.1 °C (0.2 °F)
Cold Junction Sensor Temperature Coefficient ±0.025 °/° in ambient temperature (°C or °F)
General Temperature Accuracy
±(0.008 % of mV setting + 0.006 mV) ± 0.1 °C (0.2 °F)
SOURCE Thermocouple Specifications:
Output Range -13.000 to +80.000 mV
±5 µV pp from 0.1 Hz to 10 Hz
Source Current < 8 mA
READ Thermocouple Specifications:
< ±1 LSD from 0.1 Hz to 10 Hz
Input Impedance > 1 MΩ
Comentários a estes Manuais