mA OUT, % OUT (Percent of 4 to 20 mA)
Choose this function to provide an output from 0.00 to 24.00 milliamps. The compliance voltage is a
nominal 24 VDC to provide the driving power to your milliamp receivers.
1) Disconnect one or both input wires from
the device to be calibrated.
2) Select “mA” or “% 4 to 20mA” with slide
switch
q.
3) Select “SOURCE” using slide switch w.
4) Connect the output leads of the PIECAL
334Plus to the inputs of the device being
calibrated, making sure to check polarity.
Red lead to the plus (+) input and black
lead to the minus (-) input.
The output is adjusted in 0.001 mA (0.01%)
increments by turning the knob
r while the
EZ-CHECK™ switch
e is in the “DIAL
position, or the current can be set at the
fixed points of 4.000mA (0.00%) or
20.000mA (100.00%) with switch e. Press
and turn the knob for faster dialing with
0.100 (1.00%) increments.
Start automatic stepping or ramping by pressing and holding the
r DIAL KNOB for 3 seconds. The
word STEPPING or RAMPING will flash on the display anytime the selected automatic function is
running. The 334Plus will automatically step or ramp between 4mA & 20mA (or 20mA & 4mA) for 30
seconds then reverse direction. Stop the stepping or ramping by pressing or turning the knob or
moving any switch.
Sourcing Milliamps









Using Automatic STEP & RAMP output
Move q POWER SWITCH to “mA, or “% 4 to 20 mA and
the w SOURCE / READ / 2 WIRE SWITCH to “SOURCE” or “2 WIRE”.
The starting point for automatic stepping & ramping is based
on the
e EZ-CHECK SWITCH. Place the EZ-CHECK switch
at 20 mA and the automatic stepping or ramping will start at
20.000 mA (100.00%) and the values will decrease. Place the
EZ-CHECK switch at 4 mA and the automatic stepping or
ramping will start at 4.000 mA (0.00%) and the values will
increase.
Start automatic stepping or ramping by pressing and holding the
r DIAL KNOB for 3 seconds. The
word STEPPING or RAMPING will flash on the display anytime the selected automatic function is
running. The 334Plus will automatically step or ramp between 4mA & 20mA (or 20mA & 4mA) for 30
seconds then reverse direction. Stop the stepping or ramping by pressing or turning the knob or
moving any switch.
Using Automatic Stepping & Ramping
STEPPING OUT
12.000 mA
Vista de página 3